4-3-33 Operator Panel Test

4-3-34 Peripheral checks

4-3-34-1 Printer checks

4-3-35 Turn OFF Power to LOGIQ V3/V5/V5 Expert

4-3-36 Mechanical Functions Checks

4-3-36-1 Operator Panel Movement
LOGIQ V3/V5/V5 Expert – Basic Service Manual
5726264-100 English Rev.8
The Operator Panel is tested when the LOGIQ V3/V5/V5
Expert is powered up as part of the start-up scripts, run at
every start-up.
For more info, please refer to
on page
7-1.
Check that peripherals work as described below:
Table 4-3:
Step
Task to do
1.
Press FREEZE
2.
Press PRINT on the
Control Panel
3.
Connect with Foot
Switch on USB port
and press once.
The internal printer is controlled from the Print and Store keys
on the LOGIQ V3/V5/V5 Expert's Operator Panel.
The factory default is:
Print for the standard printer
Store for the screen capture to clipboard
See 4-2-3 'Power off' on page 4-8 for more information.
Please refer to:
6-2 'LCD Monitor Adjustments' on page 6-3
Functional checks
'Diagnostics/Troubleshooting'
Peripheral Checks
Expected Result(s)
Stop image acquisition.
The image displayed on the screen is
printed on B&W printer.
To start image acquisition (the same
function as FREEZE key).