GE LOGIQ P6 Series Service Manual page 206

Table of Contents

Advertisement

GE H
EALTHCARE
5394152, Revision 5
D
IRECTION
7-7-4-3
PI128 II (LOGIQ™ P6/P6 Pro)
These programs are provided for testing the PI128 II assy.
NOTE:
A Convex Probe is needed for PI128 II Channel Test.
Table 7-4
Menu
Full Test
Assy Revision
Test
FEBUS Test
SHV Test
OPEN Test
PROBE CODE
7-7-4-4
CWD64 (LOGIQ™ P6/P6 Pro)
These programs are provided for testing the CWD64 assy.
Table 7-5
Menu
Full Test
Assy Revision
TEst
USC Interface
Test
FEBUS Test
IQ Data Read
CWD TPG Test
Probe
Connection
Channel Test
Page 7-22
FRU Test, PI128 II Diagnostics Menu
Performs all of the PI128II assy tests listed below
Test EEPROM.
Reads the assy revision of the PI128 II assy.
Checks if the DSP in SYSCONML can access the PI128 II assy via FEBUS.
Send the command to SHV resister in PI128 II assy to read SHV status.
Checks whether probe connected or not on PI128 II assy.
Checks if the 8 bits of the Probe Code in the probe connector is properly connected.
If connected probe is available, display code and Probe name.
FRU Test, CWD64 Diagnostics Menu
Performs all of the CWD64 board test listed below
Test EEPROM.Send test pattern to write to EEPROM and readDisplays the board
revision of the CWD64 assy.
Checks if the DSP in SYSCONML can access the CWD64 assy via FEBUS.
Allows SYSCONML to access FEBUS interface in the CWD64 in order to check if FE
bus line is normal.
Send the command to perform CW scanning, read IQ data from assigned memory,
analyze the data with FFT algorithm.
Read write the TPG Register in CWD64 via FEBUS.
Checks whether Dedicated CWD probe connected or not.
Generate test pattern, perform 48 channel scanning in CWD64, read the data and
analyze it.
Section 7-1 - Overview
LOGIQ™ P6/P6 P
Descriptions
Descriptions
S
M
RO
ERVICE
ANUAL
User
Level
Tools
All
None
All
None
All
None
All
None
All
Probe
All
User
Level
Tools
All
None
All
None
All
None
All
All
All
All

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents