Note: All specifications are nominal and are subject to change without notice.
Device Insulation Class
Applied Parts (Light Pad)
Input Voltage
Input Current
Leakage Touch Current
Temperature
Humidity
Atmospheric pressure
Light Box
Light Pad
Transportation Conditions & Storage Conditions (while sealed in shipping carton)
Temperature
Temperature
Temperature
Electrical Specifications
Class II Double Insulated
Type BF
~
100 – 240 V
, 50/60 Hz
0.25A – 0.75A
< 100 μA in normal condition
< 500 μA in single fault condition
Environmental Operating Conditions
+5°C to +35°C (41°F to 95°F).
15% to 90% RH, non-condensing, but not requiring a water
vapor partial pressure > 5 kPa
70 kPa to 106 kPa (10,000 ft. to -1,250 ft.)
IP21
IPX4
-25°C to +5°C (-13°F to 41°F) without humidity control.
+5°C to +35°C (41°F to 95°F) up to 90% RH,
non-condensing
+35°C to +70°C (95°F to 158°F) at a water vapor
pressure up to 5 kPa
Chapter 7
Specifications
7-1