Air Filter Cleaning/Replacement; Factors That Affect Product Life - GE BiliSoft 2.0 Operation Manual

Phototherapy system
Table of Contents

Advertisement

Air Filter Cleaning/Replacement

The air filter on the rear of the light box should be kept clean and free of lint to prevent overheating.
The filter can be removed for cleaning without needing a tool.
Figure

Factors that affect product life

Factor
Thermal stress
Light Pad folding or excessive
bending
Push up on base of Filter
Plate until tab is exposed
Description
Operating without the cooling fan running or
operating with clogged filters will reduce LED life.
Do not fold or bend the light pad or the pad
cable any tighter than 45 degrees.
Procedure
1.
Disconnect power to the unit.
2.
To remove the Air Filter, flex the Filter Cover and
remove the tabs from the lower and upper slots
that secure the Filter Cover.
3.
Clean by rinsing in water or replace if
necessary.
4.
Pat dry the filter.
5.
Once the filter is dry, place filter on back of unit
and re-install Filter Plate
6.
Insert tab in upper slot while flexing the plate
and inserting the lower tab into the lower slot.
4-5