6.4 SPU status request (item QS)

The servo status request (item QS) consists of four bytes. Bytes 0, 1 and 2 are bit coded with information
about the level alarms, operational mode and general status. Byte 3 is an ASCII character that indicates the
active operational command.
Status byte 0:
bit 0 : low level alarm
1 : low low level alarm
2 : high level alarm
3 : high high level alarm
4 : motor limit switch low
5 : motor limit switch high
6 : 1
7 : 0
Status byte 2:
bit 0 : Active interface bit 0
1 : Active interface bit 1
2 : dipped level
3 : 0
4 : general fail indication
5 : no previous ST, WD or SD command
6 : 1
7 : 0
Status byte 3:
- :
no command active
A :
balance test
B :
Block
C :
calibrate
D :
go down
F :
freeze
Page 48
Status byte 1:
bit 0 : displacer movement down
1 : displacer movement up
2 : on level
3 : test flag
4 : calibration test successful
5 : calibration test failed
6 : 1
7 : 0
Active interface:
bit 1 bit 0 Mode
0
0
0
1
1
0
1
1
N :
interface profile
L :
lock test
M :
measure frequency
R :
tank profile
T :
test gauge
U :
go up