Configuration lock key switch is displayed on the CIU 888 display (see
FIGURE 1-2).
CIU 888 display: Configuration settings are unlocked
1.2 Verifying the W&M relevant configurations settings of the CIU 888
It is recommended to verify the W&M relevant configuration settings of
the CIU 888. The purpose of verification is to establish that all W&M
relevant configuration settings stated in the certificate issued by the
legal metrology authority (e.g. NMi) have been configured correctly. The
W&M configuration settings can be verified using the CIU 888 Web
interface.
Perform the following steps to verify the W&M relevant configuration
settings:
1. Connect a PC/laptop with a standard web browser, i.e. Internet
Explorer 9 or 10, to the Service port of the CIU 888 using a Cat. 5
Ethernet cable (patch or crossover). The Service port is located at
the front of the CIU 888 (see FIGURE 1-3).
CIU 888: Service port
Sealing Guide
CIU 888
Part No.: 4417595_Rev10