1.3 Sealing the Entis Pro system connected to the CIU 888
Once the W&M relevant configuration settings have been verified, the
next step is to seal the Entis Pro system connected to the CIU 888.
Sealing the Entis Pro system is required in order to proof that the
software has not been changed.
Perform the following steps to seal the Entis Pro system:
1. Launch Entis Pro.
2. Log in using the W&M accredited for verification user account.
3. Set the following Entis Pro tasks to W&M approved:
Task Name
AboutTasks.exe
AboutTasks_ENU.dll
CiuPlusLinks.exe
CiuPlusLinks_ENU.dll
CrystalReports
DataModel_ENU.dll
Kernel
LabelControl
ObjectiveGrid
ProEssentials
ReportPrinting.exe
ReportPrinting_ENU.dll
TankDetail.exe
TankDetail_ENU.dll
TankDetail -P.rpt for English language (This template must be
imported first via About Tasks - Reports... - Import)
TankDetails NL -P.rpt for Dutch language (This template must be
imported first via About Tasks - Reports... - Import)
4. Seal the Entis Pro system.
Refer to the chapter 'About Tasks' in the Configuration manual Entis Pro
version 2.70x for Windows 7 (Part No. 4416381) for detailed information
about sealing the Entis Pro system.
1.4 Calibrating the entities of the tanks associated with the CIU 888
Once the Entis Pro system connected to the CIU 888 has been sealed,
the next step is to calibrate the entities of the tanks associated with the
CIU 888. If a tank is calibrated, it means that the tank has been
accepted by the W&M authorities for transfer.
Perform the following steps to calibrate the entities of the tanks
associated with the CIU 888:
Sealing Guide
CIU 888
5