NOTE: Ambient temperature scan is associated with the gauge associated to the first
6
Table of Contents
1. Launch Entis Pro.
2. Log in using the W&M accredited for verification user account.
3. Calibrate the required entities of each tank associated with the
CIU 888.
Refer to the section 'How to calibrate' in the Users guide Entis Pro
version 2.70x for Windows 7 (Part No. 4416380) for detailed information
about calibrating tank entities.
1.4.1 Ambient temperature calibration through CIU 888 web interface
CIU 888 web interface provides option to calibrate/uncalibrate ambient
temperature. If ambient temperature needs to be calibrated then use
this option.
1. Login into CIU 888 web page using "ciuadmin" user account.
2. Select CONFIGURE tab.
3. On siteelement tree, select the gauge that is associated with the
ambient temperature scan.
tank created in the configuration tool (CIU Service tool/Ensite Pro).
4. For ambient temperature calibration, select "Trigger calibration
command" for "Calibrate Ambient Temperature" and click Update.
Sealing Guide
CIU 888
Part No.: 4417595_Rev10