Cleaning the Electrodes,
Leadwires,
and Patient Data Cable
Clean the electrodes, leadwires, and patient data cable with a soft
cloth moistened with soapy water. You also can use a disinfectant or
cleaning agent such as CetylcideQ,
Cidex@, Lysol@, Wescodyne@,
or Ves-phene@. Wring any excess moisture from the cloth before
cleaning.
Caution
n
Do not autoclave cable or use ultrasonic cleaners.
n
The leadwires and patient module cable are not immersible.
n
Do not use abrasive materials to clean metal surfaces-scratches
on
electrodes cause artifacts.
Safety Tests
The safety tests listed below are performed at the time of
manufacture to assure compliance with these standards: IEC 601-1,
IEC 62D (CO)17, UL 544, and CSA 22.2 No. 125.
n
Chassis-to-ground resistance
n
Ground wire leakage current
H Enclosure leaka.ge current
n
Patient lead leakage current (source leakage) to ground
n
Leakage current between patient leads
1 Patient lead leakage current (sink current) with line voltage applied
Perform these tests as part of your preventive maintenance program,
and after any corrective maintenance to assure compliance with the
named standards.
To perform these tests, use a safety tester or analyzer designed
specifically for this purpose. Follow the manufacturer's operating
instructions.
Performance
Verification
and Maintenance
4-13