HP PageWriter XL M1705B Service Manual page 138

Cardiographs, mobile cart, telecommunications package
Table of Contents

Advertisement

Table 6-5. Printer
Symptom
Cause
Corrective Action
Paper doesn't move; printer
Paper out or jammed.
Replace paper or clear paper jam.
error message displayed Check
paper supply.
Defective motor/sensor
cable
Replace motor/sensor
cable assembly.
assembly.
Defective motor assembly.
Replace motor assembly.
Defective motor drive assembly.
Replace CPU assembly.
Defective roller.
Replace printer door assembly.
Paper moves then stops and
lisplays an error message.
Check paper supply.
Paper loaded incorrectly.
Dirty sensor lens.
Defective sensor.
Defective motor/sensor
cable
assembly.
Defective sensor circuit.
Defective roller.
Make sure cardiograph is loaded
properly with approved thermal paper.
Clean sensor lens.
Replace sensor assembly.
Replace motor/sensor
cable assembly.
Replace CPU assembly.
Replace printer door assembly.
Paper stops in the wrong place Wrong type of paper.
Lfter pressing
Page Advance
or
Auto). No error message.
Dirty sensor lens.
Defective sensor.
Defective motor/sensor
cable
assembly.
Defective sensor circuit.
Make sure cardiograph is loaded
properly with approved thermal paper.
Clean sensor lens.
Replace sensor assembly.
Replace motor/sensor
cable assembly.
Replace CPU assembly.
Wavy diagonal lines in printer
Improper meshing of gears in motor
Replace motor or door assembly.
self-test or distortion
of
assembly or door assembly.
lrintouts
in the time axis.
\4essage: Printer
door
2pen.
Door is ajar.
Printhead bail or leaf spring
touching contact post.
Defective sensor assembly.
Defective motor/sensor
cable
assembly.
Defective door-detect circuit.
Close door.
Check leaf spring behind printhead and
contact post.
Replace sensor assembly.
Replace motor/sensor
cable assembly.
Replace CPU assembly.
Troubleshooting
6-15