Configuration - HP PageWriter XL M1705B Service Manual

Cardiographs, mobile cart, telecommunications package
Table of Contents

Advertisement

a
Configuration
Note
,I
Yir
Configuring
the M1700A
and M1701A
Note
:'
t?
The cardiograph is designed to provide custom operation. The
configuration parameters allow you to program the cardiograph for
the most useful and efficient operation. Typically the cardiograph is
configured only during initial installation.
Software revisions A.02.00 (M1755A) or later allow you to select
either XLi, Preview Plus, or SAECG as the default configuration.
Set the Power on application global parameter to the application you
want the cardiograph to start up in.
This section describes the menu structure that enables you to
configure the cardiograph's operating parameters. Each menu choice
is described briefly in a list, then in more detail in the narrative
section that follows the list.
Press CMenu) until the following display appears:
Transmit
Store
Config
CheckDisk
Files
The Transmit
key appears only on the M1700A. Only the M1700A
has transmission capability.
Press the Config
softkey to display the main configuration menu:
Global
Configuration
ID
Transmit
Files
Exit
Note
:'
t?
If this is not the initial configuration, you must type in the
cardiograph's password (if one has been set) before you can make any
changes to the configuration.
Installation
and Configuration
2-17