A.4 SPM Protocol Command Parameter Format
Table A-8 lists the format of the command parameters as listed in the protocol document. The parameters are
listed after each command in
Parameter
Alarm Flag
Concentration
Current Loop Drive
Date
Fault Number
Format Code
Honeywell Gas No
Revision, Major/Minor
SPM Serial Number
Time
TWA (Time Weighted Average)
SPM Technical Handbook
Table
A-4.
No of Bytes
In sequential monitoring, the flag may be:
0 = Concentration only
1
1 = Level 1 Alarm
2 = Level 2 Alarm
3 = Concentration > full scale
2
16-bit binary value (0-65,535) indicating the computed concentration.
1
Proportional drive that may be on the 4-20mA current loop.
Year (7), Month (4), Day (5);
2
Year is based from 1980; 1989 would be 89 – 80 = 9;
9 = 0 0 0 1 0 0 1
A number indicating a specific type of fault. Consult SPM manual for fault
1
codes.
Top bit indicates ppm or ppb (1= ppm). The remaining 7 bits indicate decimal
1
position. Therefore, a parameter of (Hex) 81 (1000 0001) would have
concentration indicated as xx.x ppm.
1
Defines gas being monitored.
1
Software revision level of the SPM EPROM installed in the instrument.
2
Serial number of the SPM instrument.
2
Hours (5), Minutes (6), Seconds/2 (5)
16-bit binary value (0 to 65,535) representing the 8 hour TWA concentration
2
value.
Table A-8: SPM Protocol Command Parameter Format
SPM Single Point Monitor
Parameter Description/Components (no of bits)
A-7