Equipment setup
Leadwire installation
3-4
NOTE
When new batteries are installed, all LEDs on the transmitter flash, then
flash again twice to acknowledge the new battery installation. The flashing
LEDs do not indicate good leads. You must press the Verify Leads button to
check lead status.
When the Change Battery LED starts flashing, the transmitter has
approximately one hour of reserve power before the unit shuts down.
When a low battery condition occurs, approximately two hours before the
battery loses power, a Low Battery message displays at the CIC Pro center
without an audible alarm. When the battery is dead, an audible system
warning alarm is generated at the CIC Pro center.
For optimum performance, follow these guidelines:
Install two new AA alkaline batteries when you begin monitoring a new patient.
Install two new AA alkaline batteries when the Change Battery LED flashes.
Do not use rechargeable batteries.
To install batteries in the transmitter, follow these steps.
1.
Locate the battery cover at the bottom of the transmitter.
2.
Slide the cover over to open the battery compartment.
3.
Insert the batteries according to the polarity signs on the lower back side of the
transmitter.
4.
Close the battery cover.
The transmitter can use the following Multi-Link leadwire sets:
Multi-Link 6-leadwire set
Multi-Link 5-leadwire set
Multi-Link 3-leadwire set
To install a leadwire set into the transmitter, align the leadwire pins with the
connector on the top of the transmitter, then push the leadwire set firmly into the
transmitter.