Low Voltage Unit (Fru No. 703) - GE LOGIQ 400 Service Manual

Hide thumbs Also See for LOGIQ 400:
Table of Contents

Advertisement

GE MEDICAL SYSTEMS
REV 9
6–2–87 Low Voltage Unit (FRU No. 703)
Time Required
30 Minutes
Tool Required
Screwdriver
Procedure
Refer to ILLUSTRATION 6–122 on page 6–219.
1.
Turn OFF the system and unplug the main power cord.
2.
Remove the Side Cover Left (FRU 301). Refer to 6–2–31 on page 6–127.
Unscrew four screws ( 1 – 4 ).
3.
4.
Remove the Left PS Cover.
Unscrew two screws ( 5 and 6 ).
5.
6.
Pull the Low Voltage Unit outwards.
Unscrew fifteen screws ( 7 –
7.
8.
Pull the Low Voltage Unit out.
9.
Remove the Plastic Cover as shown in ILLUSTRATION 6–122 on page 6–219.
10. Disconnect one connector (
11. Unscrew two screws (
12. Remove the Low Voltage Unit.
21
) to remove wires.
22
).
23
24
and
) to remove the FG wires.
6–218
LOGIQ 400 SERVICE MANUAL
2127661