Force Bump; Data Logging; Ble; Assigned - Honeywell BW Solo User Manual

Hide thumbs Also See for BW Solo:
Table of Contents

Advertisement

For Manual Calibration: At the "Cal Now" screen, press and hold the button to begin the process.
Follow the calibration instructions. Once the Calibration has passed, the user may proceed to the
idle screen. If calibration fails, then the instrument will revert to the "Cal Now" screen.

9.10.1.2.Force Bump

When this feature is enabled, the display reads "Bump Now." The user cannot pass this screen
without bump testing either manually or through IntelliDoX before proceeding to Idle Screen.
For Manual Calibraion: Press and hold the button to start the bump process. Follow the
instructions for the bump process. Once the bump test has passed, the user may proceed to the
idle screen. If the bump test fails, then the instrument will revert to the "Bump Now" screen. If no
action is taken, then the unit will automatically shut off after 60 seconds.
The default setting for bump and calibration reminders is off.

9.11. Data Logging

Turn datalogging on or off and set the frequency interval (in seconds) for data collection.
Note: Turning data logging off hides the "Interval" setting option. The default setting for "Interval" is
005 seconds, and the interval can be set between 001 and 300 seconds.
The default setting for data logging is on.

9.12. BLE

On instruments that feature BLE (Bluetooth Low Energy), turn BLE communication on or off, so that
the instrument is discoverable or not discoverable.
Note: If BLE operation is desired, this option must be specified at time of order.
• When BLE is turned off, the BLE "off" icon is displayed.
• When BLE is turned on, the BLE "on" icon is displayed.
The default setting for BLE is on.

9.13. Assigned

The instrument can be assigned to a person, and their name can be programmed into it. You can
also turn assignment on or off, or edit/rename an assignment. The name can be up to 10
characters long.
The following are available for entry:
• Digits 0 to 9
• Upper-case characters A to Z
• A space

9.14. Zone

A zone's name can be programmed. You can turn location on or off, and name, edit, or rename it.
The following are available for entry:
Honeywell BW Solo User Manual