7.2 PM 620 modules

Operating temperature
Storage temperature
Ingress Protection
Humidity
0 to 30°C (32 to 86°F):
30 to 50°C (86 to 122°F):
Shock/Vibration
EMC
Electrical safety
Pressure safety
Approved
Size
Weight
Pressure connections

7.2.1 Pressure data (PM 620 modules)

[EN] English - K0457
For a full specification of the PM 620 module, refer to the
datasheet supplied on the CD (CD: P/N UD-0002)
Table 7-3: General specification (PM 620
-10 to 50°C (14 to 122°F)
Calibrated range: 0 to 50°C (32 to 122°F)
-20 to 70°C (-4 to 158°F)
IP65
Relative humidity (RH)
0 to 95% RH non-condensing
0 to 60% RH non-condensing
Def Stan 66-31, 8.4 cat III
Electromagnetic compatibility: BS EN 61326-1:2006
Electrical - BS EN 61010:2001
Pressure Equipment Directive - Class: Sound Engineering Practice
(SEP)
CE Marked
Length, Diameter ≈ 58 x 44 (2.3 x 1.7 in)
Maximum: ≈ 0.11 kg (0.2 lb)
Positive (male) port: Attaches to PV 62x pressure stations only
M5 (female) reference pressure port.
Table 7-4: Maximum working pressure (PM 620 modules)
Ranges:
≤ 350 mbar (5 psi)
> 350 mbar (5 psi)
Issue 1
)
MWP
2 x FS
1.25 x FS
Specification 7-3