5.3.2 DPI 620 calibrator: Set the pressure function

1
TAP
Select Function
5-4 Pressure calibrator operation (DPI 620)
If necessary, change the Automation options for the CH1
and/or CH2 output functions: item
This includes: Nudge, Span Check, Percent Step,
Defined Step, Ramp; refer to the user manual - K0449
If necessary, change the Configuration: item
This includes: Data Logging, Documenting, and Advanced
options; refer to the user manual - K0449
When all the software selections are complete, make the
applicable pressure and electrical connections. Examples:
Gauge/indicator calibration (Section 5.3.6); Transmitter
calibration (Section 5.3.7); Switch test (Section 5.3.8)
This example shows the sequence to set the pressure function.
It is a similar procedure for other functions; refer to the user
manual - K0449.