Pm 620 Module Parts; Assembly Instructions - GE Druck PV 622 User Manual

Druck pv 62x series. pneumatic/hydraulic pressure stations
Table of Contents

Advertisement

Issue 1
5.2.2 PM 620 module
parts
1

5.2.3 Assembly instructions

1
4
3
5
5-2 Pressure calibrator operation (DPI 620)
Caution: To prevent damage to the PM 620 module, only
use it within the specified pressure limit on the label.
1.
Pressure module (PM 620) with a pressure connection,
reference port (a) and a label. The label includes:
a
Pressure range. Example: 20 bar g (g: gauge; a: absolute);
serial number (S/N); manufacturer: name, address, website
Table 5-1: Recommended pressure modules (PM 620)
PV 62x model
PV 621 models
PV 622 models
PV 623 models
Step
1.
2.
3.
2
4.
5.
PM 620 module range
25 mbar to 20 bar (0.36 to 300 psi)
25 mbar to 100 bar (0.36 to 1500 psi)
70 to 1000 bar (1000 to 15000 psi)
Procedure
Lower the DPI 620 calibrator into the moulded
compartment.
Press on the bottom end of the calibrator until it
latches in position.
Attach a PM 620 module with the correct range
and type.
Tighten it until it is hand tight only.
When this symbol flashes at the top of the display,
it shows there is communication between the
module and the calibrator.
If applicable, attach one of the optional PRVs; see
Chapter 1 (Instrument parts, accessories
and options).