Example Procedure: Gauge/Indicator Calibration - GE Druck PV 622 User Manual

Druck pv 62x series. pneumatic/hydraulic pressure stations
Table of Contents

Advertisement

TAP the window
(Maximise)
1

5.3.6 Example procedure: Gauge/indicator calibration

1
2
[EN] English - K0457
Select Settings
2
See the new reading and go back to the Home
window.
3
1. Assemble the pressure calibrator with the correct PM 620
module and PRV (if applicable); see Section 5.2.3.
2. Set the applicable software options; see Section 5.3.1
(Procedure overview). This example shows one function:
Pressure (P1) is set to measure the pressure supplied by
the pressure station.
3. To attach the device under test, use the specified
procedures for your PV 62x pressure station*.
Select Zero
Accept the adjustment value
3
5
Pressure calibrator operation (DPI 620) 5-7
Issue 1
4
4