Spo2 (Perform Only If Equipped With Spo2 Module) - GE CareScape V100 Service Manual

Vital sings monitor
Hide thumbs Also See for CareScape V100:
Table of Contents

Advertisement

SpO
(perform only if equipped with SpO
2
2048723-002C
Maintenance: Calibration procedures and tests
1.
Turn on the verifier device, using either a 9-volt battery or the power supply.
Make sure the red LED is illuminated.
Power On LED
Power supply
jack
2.
Allow device approximately 5 minutes for warm-up and stabilization time.
3.
Allow certified master and the instrument to be tested to acclimate in the
same ambient temperature for at least 10 minutes.
4.
For all instruments, make sure the lens at the tip of the probe is clean. To
clean, use an alcohol prep or a swab dipped in alcohol, followed by a damp
wipe with water to remove any residue.
5.
Alternately insert the reference instrument and the instrument being verified
into the aperture opening, comparing the readings.
6.
Record and verify the difference between the temperature reading of the
CM and the instrument being verified is ±0.2°C (0.4°F).
If the readings differ by more than the acceptable field limits, repeat this
procedure. If the readings still differ by more than the acceptable limits, the
device fails.
7.
Confirm test results are recorded on the
1.
Connect the appropriate SpO
the probe on your finger.
2.
Verify the unit displays a:
Pulse value
Saturation value
Signal Strength Bar Graph
CARESCAPE V100 Vital Signs Monitor
Certified master (CM) or
instrument to be verified
"Test results form"
module)
2
probe and cable to the SpO
2
On/Off switch
Battery
compartment
Portable black
body verifier
on page 4-29.
connector. Place
2
4-21