Error Code
Reply
Seriousness
Keys
-305
OK
Error
-306
OK
Error
-310
OK
Error
Error Code List for Version 6.00
[Operation after Reply Key is touched]
#1:Error
#2:Calibration Error
#3:Collimator may be used.
[Cause]
This error is not currently used.
[Operation after Reply Key is touched]
Perform imaging again.
#1:Error
#2:QA Process Error
#3:Image Analysis Error. Adjust image process parameter
manually.
[Cause]
Feature extraction may have failed.
[Operation after Reply Key is touched]
The user must process the image manually.
#1:Error
#2:REX Value Warning
#2:REX Value Warning
#3:REX is 999. It is under the minimum limitation value
9999.
Please reconfirm the generator setting or exposure
condition.
If the condition is correct, adjust the image processing.
[Cause]
This error occurs when the REX value acquired by image
processing after the imaging is less than the limitation value.
[Operation after Reply Key is touched]
Nothing
[Note]
999: REX value obtained by image processing
9999: lower limit value
- 10 -