3 The measurement is completed in four minutes, and coefficient (calculated value)
will be displayed.
[Standard value of the viscometer calibration] 1200 to 2200
*The value stated above is a reference. There will be no problem even if the coefficient
value displayed is other than stated above as long as data of Plunger falling time keeps
within a similar range.
Viscometer calibration
<the previous calibration>
Exec.time 2018/07/07 11:30
Plunger falling time
X.XX (sec)
1st
2nd
X.XX (sec)
3rd
X.XX (sec)
4th
X.XX (sec)
5th
X.XX (sec)
Average
X.XX (sec)
Inside temperature XX ( )
Standard
X. XXX (mPa▪s)
Coefficient
XXXX
When updating coefficient, press [OK].
If "Fault" is displayed for "Proc. Status", follow the instructions given in
"4.12 Troubleshooting when Fault Occurs in Viscometer."
4 Press OK.
If the OK key is pressed when "Proc. Status" is "Ended", the coefficient will be
registered, and the circulation control screen will be displayed.
[Note]
The coefficient value will not be registered when Cancel is pressed instead of OK key.

2.6.2 Operation test of viscometer

●This function allows you to check the viscometer operation and measure the current ink
viscosity. Note that this operation test is not to determine the coefficient.
● It is not necessary to use new ink for operation test.
[Stop
]
< the current calibration >
Exec.time 2018/07/07 12:45
Plunger falling time
X.XX (sec)
1st
2nd
X.XX (sec)
3rd
X.XX (sec)
4th
X.XX (sec)
5th
X.XX (sec)
Average
X.XX (sec)
Inside temperature XX ( )
Standard
X. XXX (mPa▪s)
Coefficient
XXXX
Ink viscosity
XXX
Proc. Status: Ended
2-25
2018.07.07 12:45
Com=0
HOME
Cancel
OK
Coefficient
OK