ELECTRICAL CONNECTIONS
3

3.5.4 Connection diagram for measuring sensor, field housing

DANGER!
The device must be grounded in accordance with regulations in order to protect personnel
against electric shocks.
• If a shielded field current cable is used, the shield must NOT
the signal converter.
• The outer shield of signal cable A or B in the signal converter housing is connected via the
strain relief terminal.
• Bending radius of signal and field current cable: ≥ 50 mm / 2"
• The following illustration is schematic. The positions of the electrical connection terminals
may vary depending on the housing version.
Figure 3-17: Connection diagram for measuring sensor, field housing
1 Electrical terminal compartment in housing of the signal converter for signal and field current cable.
2 Signal cable A
3 Signal cable B
4 Field current cable C
5 Connection box of measuring sensor
6 Functional ground FE
36
www.honeywell.com/ps
NOT be connected in the housing of
NOT
NOT
34-VF-25-53 iss.3 GLO Aug 10 US