Sets and queries the save/recall register number to alternate the
foreground state of the instrument. The *RST value is 1.
SYSTem:ALTernate:STATe ONlOFFlllO
l
SYSTem:ALTernate:STATe?
l
Sets and queries the state of the Alternate State function.
l
Changes the GPIB's (General Purpose Interface Bus) address.
The *RST value is 19.
SYSTem:DUMP:PRINter?
l
Causes a dump of the display contents to be made to the HP-IB.
l
Returns the next message in the error queue. The format of the
response is : <error number>,<error string>
where the error number is as shown in the "Error Messages" section
and error string is :
"<Generic HP-SL string>;<More specific information>"
An example response to SYST:ERR? is
-23, "NUMERIC OVERFLOW ;YOU PUT IN A NUMBER TOO BIG"
SYSTem:KEY:ASSign <num>,<num>
l
This assigns the first numeric value (key code) to the second numeric
value (user menu key number). Every softkey menu item is given a
unique key code (as if it were a dedicated front panel key). Refer to
the entry SCPI KEY NUMBERS for a listing of key codes.
SYSTem:KEY:CLEar <num>lALL
l
Clears the user menu. If <r-mm> is sent, only that particular key
number is cleared, If ALL is sent, all menu keys are cleared.
SYSTem:KEY :DISable SAVE
l
The save key grouping is disabled. In our box this disables the save
state feature. (Save Lock)
Unlocks the save registers. The *RST value is for the save registers
to be enabled.
SYSTem:LANGuage SCPIlCIILlCOMPatible
l
Causes the synthesizer to perform a language switch to another
language system.
Operating and Programming Reference S-53