8-9.
Test Menu Utility Softkey
Description
The test menu [Utility] softkey gives you
the capability
of:
Changing the synthesizer's defined
option configuration (option configuration defined by the
cal
constants) from your computer,
Displaying data from previously
run
tests, or
Displaying
the test status of
previously tested instruments.
Use the change
option feature to simplify instrument reconfiguration after you add or delete
an
instrument option.
Test Equipment
None Required.
Procedure
1.
After
you
select
the test
menu [Utility] softkey, the program displays
the
following softkey
choices:
[Chg Option]
lets you
change the synthesizer's defined option configuration.
After
you select this softkey, input
each
of the
synthesizer's three-digit option strings (one
at
a
time), pressing
[ENTER] after
each entry. After you have input the options, check the option
list.
When
the option entry
is
complete,, the program automatically updates the calibration constants
in
the synthesizer, and the data
file
on the data disc, to reflect the option configuration you
entered.
[Dsp Data]
lets you
display or print the
test
data from
previously run tests (see the
NOTE
below).
The
data displayed
is
for the instrument
whose
serial
number
is
currently
in
the
computer's
memory, and for the type of testing
you selected while
in
the
TEST SELECTION
menu (either
the
attenuator calibration or the operation
verification).
When
you select this
softkey, the program displays the test menu
and lets you
output
all
the test
data ([ALL DATA]), output
the data for one test
([SELECT]), or
return to the
UTILITIES
menu
([EXIT]).
The
[SELECT]
softkey outputs the data for the test
at which the cursor points. Use the keyboards
UP or DOWN arrow keys or the front panel knob on the keyboard to position the cursor.
[Prt Status]
lets you
display or print the test status for the synthesizer
whose
serial number
is
currently
in
the computer's memory.
The
status displayed
is
for the type of testing you selected
while
in
the
TEST SELECTION
menu (attenuator calibration or operation verification).
[Exit]
returns
to
the
test menu.
Attenuator Calibration and Operation Verification Test