8-8.
Frequency Switching Time
Test
Description
This test test verifies that the
synthesizer meets
its
frequency switching time specification.
It
uses the
synthesizer's rear panel negative
blanking output and an HP
5316A to measure the time
it
takes the
instrument
to switch from
one CW
frequency to another.
The frequencies
tested are those that cause
the worst
case
switching time delay.
Test Equipment
Universal
Counter
..
BNC Cable (m)
-
(m)
..........
HP 5316A
HP Part No. 8120-1840
Procedure
1.
Set
up the equipment as shown on the computer
CRT.
2.
Select [CarryOn].
The [JANITOR]
softkey returns
you to the test menu.
2.
Select the desired softkey and respond to
the
program
prompts:
[FullTst] performs
the entire frequency switching time test, and compares the
measured
data to
the synthesizer's specifications. Select
this
softkey to
meet the
requirements of the
performance
test.
When
this
test completes, the test results are displayed. Select
[Carry On]
and then either
[Re-
Test]
(to return
to the frequency
switching time menu), or
[JANITOR]
(to
return to the test menu).
[Other]
lets you
select the two CW frequencies between which
the
test
will
switch. You have
a
choice
of:
[A to B] entering two
frequency
points;
[A
to
A+d]
entering one frequency point and
the
frequency step
size;
[FullTest] running the
full
test; or
[Janitor] returning
to the
test menu.
If
you select
[A to B] or [A to
A+d],
the frequency you
enter
must be
in
MHz.
[JANITOR]
returns you to the test menu.
8-12
Tests/Utilities
Attenuator Calibration and Operation Verification Test Software

Advertisement