Chapter
14.
RF Attenuator Calibration
14.0 Description
This program performs three functions:
It
adjusts the instrument's attenuator calibration constants to compensate for errors
In
the
90 dB step attenuator
(for those instruments so
equipped)
It
verifies power accuracy
at
each attenuator step
It
sums the attenuator
performance data
with output power
accuracy data and compares
the results with the
instrument's performance specifications
NOTE
This test
will
not run unless
the
IF
attenuator has been characterized per
instructions
in
Chapter
6
and the Output Power Accuracy test has been
run.
The data
files
generated
by
the
IF
attenuator and Output Power
Accuracy programs must be
on
the
data
disc currently
in
the disc drive.
14.1
Test Equipment
Local Oscillator
Spectrum Analyzer
.....-..„„
-
-
-
Mixer
„„.„„„
.
.
.
.
..
„.
.
20
dB Attenuator
.-„.-.
10
dB Attenuator
....
....
..
...
..
...
Step Attenuator (Calibrated
at
1
MHz)
Low Pass
Filter
-,
....
Amplifier
-
.
.
-
...
....
Cables:
Semi-rigid coax,
SMA (m)
to
SMA (m)
BNC (m)
to
BNC
(m),
(7 required)
-
.
Adapters:
APC 3.5 (f)
-
APC 3.5
(f)1
-
-
......
..
APC 3.5
(f) to
Type
N
(m)2
..
..
..
.
SMA
(m) to BNC (f)
-
-
-
-
-
1
Required
tor an
HP
8340A/B OUT
only
2
Required
for an
HP
8341A/B OUT
only
HP 8340A Option
H01
or
HP 8340A/B
Option 001
-
-
.....
HP
3585A
.
-
HP
P/N 0955-0307
HP 8493C
Option 020
-
-
HP
8493C
Option 010
-
-
-
HP
355D
HP P/N 9135-0260
.
HP
8447F
HP P/N
08340-20124
HP P/N 8120-1840
HP P/N 5061-5311
HP P/N 1250-1744
HP P/N 1250-1200
HP 8340A/B
and
HP 8341A/B
RF Attenuator Calibration
37