Meter Adrs
Function Group
Menu Map 8
Description
Programming Codes
See Also
Meter On/Off AM
Function Group
Menu Map
Description
Programming Codes
See Also
HP 8360
User's Handbook
SYSTEM
In cases where the synthesizer is capable of acting as a controller
to an HP 437B power meter, this softkey enables you to set the
programming address of the power meter. The address value can
be set from 0 to 30, with the factory default address set at 13. The
address value is stored in non-volatile memory.
SCPI:
DIAGnostics:INSTrument:PMETer:ADDRess <num>
Analyzer:
NONE
"Optimizing Synthesizer Performance," in Chapter 1.
"INSTALLATION," Chapter 3.
4
This softkey (Option 002 only) lets you display the value of the depth
of the externally-generated amplitude modulation.
SCPI:
MEASure:AM?
Analyzer:
NONE
Operating and Programming Reference M-9