Calibration; Calibration/Validation Overview - Siemens SIPROCESS GA700 Operating Manual

Continuous gas analysis, operating with the local user interface
Hide thumbs Also See for SIPROCESS GA700:
Table of Contents

Advertisement

Functions
6.1 [1] Quick Start
1. "1 Quick Start" > "4. Measuring ranges" > "n component" main menu
The menu includes four navigation lines for setting the measuring ranges and one
navigation line for setting autoranging.
2. Call autoranging: Navigation line "Autoranging".
The parameter display of autoranging has a parameter field and a display box.
3. Specify measuring range: "Measuring range selection" parameter field.
– Authorize access with Standard PIN, if necessary.
– Select one of the offered measuring ranges or autoranging:
6.1.7

[1.5] Calibration

6.1.7.1

Calibration/validation overview

Introduction
This menu enables you to make simple calibrations or validations, for example, as part of
initial commissioning. This means the offered range of functions is very limited. Automation
functions, such as the use of the calibration pool, cannot be activated using this menu.
"Factory calibrations" also are not part of the functions included in the Quick Start menu.
You can find detailed information on how to fully access all possibilities of the
calibration/validation functions of the device in the section → [2.08] Settings > Calibrations.
Difference between calibration/validation
Calibration
In a calibration, the device is adapted to its measurement task(s). For the actual value to be
equal to the setpoint, the device changes one or more calibration factors during calibration.
92
If the message "Active" is displayed in the "Autoranging" line, autoranging is activated.
If "Not possible" is displayed, adjust the start and end values for the individual
measuring ranges: Setting the measuring ranges (Page 89).
Operating with the Local User Interface
Operating Manual, 06/2017, A5E31930478-05