Siemens SIPROCESS GA700 Operating Manual page 158

Continuous gas analysis, operating with the local user interface
Hide thumbs Also See for SIPROCESS GA700:
Table of Contents

Advertisement

Functions
6.8 [2.09] Settings > Inputs/outputs
Table 6- 31
Function (pre-
selection)
Measuring ranges
Table 6- 32
Function (pre-
selection)
Process tags
Table 6- 33
Function (pre-
selection)
Solenoid valves
158
Function (pre-selection) "Measuring ranges"
Purpose/resulting settings
Specify measuring range to be displayed.
1. "Component" parameter box:
Authorize access with corresponding PIN, if necessary.
Select available component.
2. "Function" parameter box:
Select one of the four available measuring ranges.
Function (pre-selection) "Process tags"
Purpose/resulting settings
Output measured value of the used process tag.
1. "Function" parameter box:
Authorize access with corresponding PIN, if necessary.
2. Select one of the offered process tags.
Function (pre-selection) "Solenoid valves"
Purpose/resulting settings
Switch gas paths.
1. "Component" parameter box:
Authorize access with corresponding PIN, if necessary.
Select available component.
2. "Function" parameter box:
Set the following alternative gas paths:
Sample gas
Zero gas measuring ranges 1 to 4 (MR1 to MR4)
Span gas measuring ranges 1 to 4 (MR1 to MR4)
Operating with the Local User Interface
Operating Manual, 06/2017, A5E31930478-05