Functions
6.14 [2.20] Settings > Service
6.14.5

[2.20.10] Name/unit

Procedure
1. Main menu > "2 Settings" > "20. Service" > "10. Name/unit"
This menu is used for setting the component name and to select the appropriate unit.
Note
Component names with several meanings/assigned several times
If you assign names several times, the components cannot be distinguished in the main
view and during navigation.
• Assign only unique component names, for example, "O2(A)" and "O2(B)" for a device
• Make sure that you do not assign multiple names.
2. Enter the name of the measured component: "Component name" parameter field
– Authorize access with Expert PIN, if necessary.
– The component name can have a maximum of 31 characters. But only seven
– Enter the name with no more than seven characters.
3. Specify the measurement unit of the component: "Unit" parameter field.
Note
Reference temperature during conversion of units
When converting the measured values to volume-based units, e.g., vpm, mg/m³, the
reference temperature is 0 °C.
Select one of the units included in the selection list.
– Selection "ppm"
– Selection "User-defined"
200
with two OXYMAT 7 modules.
characters will be visible on the display.
When you set this unit, the conversion factor for the conversion into the unit "ppm" is
displayed.
The current offset and the measured value in the unit "ppm" are also displayed.
Additional parameter fields are displayed with this selection:
Parameter field
"User-defined unit"
"Factor"
"Offset"
Action
Assign name of unit, up to seven characters.
Enter conversion factor for conversion to "ppm".
Enter offset for conversion of user-defined unit to "ppm".