Siemens ADVIA LabCell Operating Instructions Manual page 73

Centrifuge module interface
Table of Contents

Advertisement

Centrifuge Module Interface Operating Instructions
2. Open the robot enclosure access door and place the hand crank in the
hand crank receptacle.
Figure 7:
1.
2.
3.
3. Press and hold the Z-brake release and turn the hand crank to raise the
robot assembly.
CAUTION
Do not operate the Centrifuge module with the hand crank in the
module. Operating the Centrifuge module with the hand crank in place
can cause system damage. Store the hand crank in an easily accessible
location.
4. After raising the robot assembly, remove the hand crank.
092D0001-02 Rev. A
Hand Crank Placement
Hand crank
Hand crank receptacle
Z-brake release