Siemens ADVIA LabCell Operating Instructions Manual page 48

Centrifuge module interface
Table of Contents

Advertisement

Maintenance
4. Pull the robot arm toward you until the robot is above the scale.
Figure 4:
1.
2.
5. Inspect the gripper for damage.
If the gripper is damaged, contact your local technical service provider
to determine if the gripper needs replacement.
6. Dampen a cleaning cloth with 70% isopropyl alcohol.
You can use an alcohol prep pad instead of the damp cloth.
WARNING
Do not clean the gripper without turning and holding the gripper release
switch or locking the gripper open. Releasing the switch too soon while
cleaning the gripper can cause bodily injury. Turn and hold the switch
until the robot gripper locks open, at least 10 seconds.
page 48
Robot Arm above Scale
Robot Gripper
Scale
Centrifuge Module Interface Operating Instructions
092D0001-02 Rev. A