GE Dash Port 2 Service Manual page 69

Table of Contents

Advertisement

Problem
The monitor is not properly secured to the
docking station.
The text and waveforms on the remote
display appear broken-up.
No picture appears on the remote display
and the following conditions apply:
The remote display's illuminated LED
is flashing amber.
The docking station's illuminated
communication indicator is green.
No picture appears on the remote display
and the following conditions apply:
The remote display's illuminated LED
is flashing amber.
The docking station's illuminated
communication indicator is amber.
Revision A
Troubleshooting: Troubleshooting Procedures
Problems and Solutions Table
Cause
The docking station's clamping mechanism
is not properly engaged or is defective.
The monitor's GCX plate or footpad is
either missing or not installed correctly.
The remote display's horizontal and
vertical adjustments are not synchronized
with the monitor's video output. As a result,
not all of the pixels required to make a
complete text or waveform image are being
displayed.
The remote display or the docking station
is defective.
The docking station is defective.
Dash Port 2 Docking Station
2000966-138
Solution
See
"Mount the Dash Monitor"
10.
1. Verify the security lever starts in the far
right position BEFORE placing the
monitor onto the docking station.
2. Verify the monitor is seated properly on
the docking station.
3. Move the security lever to the far left
position.
4. If the monitor cannot be properly
secured to the docking station,
the Top Housing FRU"
on page 6-18.
See the assembly drawings provided in the
Dash Patient Monitor's Service Manual to
verify that the GCX plate and footpad are
installed correctly.
1. Allow the remote display to automatically
adjust to the video signal. See
"Automatic Adjustment"
on page 3-12.
2. If the text and waveforms continue to
appear broken-up, manually adjust the
remote display's video signal. See
"Manual Adjustment"
on page 3-13.
1. Connect the docking station to a known
good remote display.
2. If the picture appears on the remote
display, replace the defective remote
display.
3. If a picture does not appear, then the
docking station's PCB must be replaced.
See,
"Replace the Connector Assembly
FRU or PCB FRU"
on page 6-19.
1. Remove the docking station and connect
a known good docking station.
2. If a picture appears on the remote
display, then the docking station's PCB
must be replaced. See,
"Replace the
Connector Assembly FRU or PCB FRU"
on page 6-19.
on page 3-
"Replace
5-11