Technical Specifications

Docking Station

Remote Display

Revision A
Appendix A – Technical Specifications: Technical Specifications
Item
Power Requirements
with monitor at 115V setting
with monitor at 230V setting
Power Consumption (fully loaded)
Communication
Network Connection
Operating Conditions
operating temperature
relative humidity, non-condensing
atmospheric pressure
Cooling Method
Storage Conditions
Physical Specifications
Certification
See
"Appendix B – Remote Display"
remote display specifications.
Dash Port 2 Docking Station
2012659-002
without monitor
88-270 VAC RMS, 47-63 HZ
Maximum input current is125 mA RMS
88-135 VAC RMS, 47-63 HZ
176-270 VAC RMS, 47-63 HZ
75 W
RJ45 connector/10BaseT.
Video Output
15 pin female D connector. 640 x 480. See
"Appendix B – Remote Display"
additional specifications.
AUX Port
RJ45 connector/RS232 (hardware interface type
autoport). To connect M-port devices to the AUX
port use the M-port to Auto-port adapter, PN
2001973-001.
0 to 40°C (32 to 104°F)
5 to 95% at 40°C (104°F)
altitude
-610 to 4,570 m (-2000 to 15, 000 ft.)
425 to 817 mmHg (56 to 109 kPa)
convection
maximum
70°C (158° F) at 95% relative humidity
minimum
-40°C (104° F) at 85% relative humidity
height
13.9 cm (5.5 in)
width
27 cm (10.5 in)
depth
20 cm (8.0 in)
weight
1.95 kg (4.3 lb.)
maximum load
6.8 kg (15 lb.)
UL 2601-1 classified
UL classified for CAN/CSA C22.2 No. 601.1
IEC 60601-1 and EN 60601-1 certified
CE Marking for 93/42/EEC concerning medical
devices
for the required and recommended
Specification
for
A-3

Advertisement

Table of Contents
loading