4-10. Replacing the Processor Assembly

4-10. Replacing the Processor Assembly
Preliminary Preparations
Open the door at the bottom of the unit.
Removal
1. Removing the camera cable
(1) Disconnect the camera cable from the clip
plate (coaxial cable and flat cable).
n
If the camera cable is bound to the handle of
the processor assembly, unbind it also.
(2) Remove the two screws, and remove the clip
plate.
(3) Loosen the four screws (with stopper), and
remove the front cover.
(4) Disconnect the flat cable from the connector
CN6027 of the MY-89 board.
(5) Remove the three coaxial cables from the
connectors CN402, CN702, and CN102 of
the ADC-36 board.
4-14
Screw (with stopper)
Front cover
Screw
(with stopper)
PSW3 x 6
Clip plate
Top view
Harness band
Notch
Clip plate
ADC-36 board
MY-89 board
Flat cable
Coaxial cable
Screw (with stopper)
Camela
cable
Hundle
CN402
CN702
CN6072
CN102