Setting Up Ports and Channels
110
How to Set Up a Port or Channel
for Generated Signals
To set the parameters for a DUT port or terminal that receives generated
signals or a channel that generates signals:
1 Open the Parameter Editor for the port or channel (see "How to Start
the Parameter Editor for Setting Port Parameters" on page 107 or
"How to Start the Parameter Editor for Setting Channel Parameters"
on page 108.
Figure 37
General Timing Parameters for a Data Generator
Note that the window displays the identification of the frontend and may
offer more parameters depending on the frontend.

How to Set Generator Timing Parameters

1 Check the units.
The Parameter Editor displays default units and has default vernier
steps. Both can be changed (see "How to Change Units and/or Vernier
Steps" on page 69).
2 Set the timing options:
– Delay: The delay is relative to the system clock. A negative delay can
be set if the system clock has been delayed.
– Width or Duty Cycle: They are mutually dependent.
If you have typed a number, terminate your input with the Enter or
Return key. This updates the other field (Width or Duty Cycle).
HP 81200 Data Generator/Analyzer Platform User Guide, Revision 2.1