Checklist - GE Dash 3000 Service Manual

Patient monitor
Hide thumbs Also See for Dash 3000:
Table of Contents

Advertisement

Checklist

Details of test Configuration, Conditions, and Test Equipment for (DUT) Device Under Test (Complete one sheet per device)
Product/Model
ID Number
Manufacturer
Step
DUT
Electrical Safety Tests
A
Ground Continuity
1.
Measured Resistance
B
Earth Leakage Current
1.
Pol Norm, Neut Clsd
2.
Pol Norm, Neut Open
3.
Pol Rev, Neut Open
4.
Pol Rev, Neut Clsd
C
Enclosure Leakage Current
1.
Pol Norm, Neut Clsd, Gnd Clsd
2.
Pol Norm, Neut Open, Gnd Clsd
3.
Pol Norm, Neut Clsd, Gnd Open
4.
Pol Rev, Neut Clsd, Gnd Open
5.
Pol Rev, Neut Open, Gnd Clsd
6.
Pol Rev, Neut Clsd, Gnd Clsd
Patient Leakage Current
D
to Ground
1.
Pol Norm, Neut Clsd, Gnd Clsd
2.
Pol Norm, Neut Open, Gnd Clsd
3.
Pol Norm, Neut Clsd, Gnd Open
4.
Pol Rev, Neut Clsd, Gnd Open
5.
Pol Rev, Neut Open, Gnd Clsd
6.
Pol Rev, Neut Clsd, Gnd Clsd
Patient Leakage Current
E.
Mains on Applied Part
1.
Pol Norm, Neut Clsd, Gnd Clsd
2.
Pol Rev, Neut Clsd, Gnd Clsd
BISx Patient Leakage Current
F.
to Ground (option)
1.
Pol Norm, Neut Clsd, Gnd Clsd
2.
Pol Norm, Neut Open, Gnd Clsd
3.
Pol Norm, Neut Clsd, Gnd Open
4.
Pol Rev, Neut Clsd, Gnd Open
5.
Pol Rev, Neut Open, Gnd Clsd
6.
Pol Rev, Neut Clsd, Gnd Clsd
BISx Patient Leakage Current
G.
Mains on Applied Part (option)
1.
Pol Norm, Neut Clsd, Gnd Clsd
2.
Pol Rev, Neut Clsd, Gnd Clsd
D-2
Checklist: Checklist
Customer Asset Tag
Prior to testing verify all equipment is calibrated via "Cal" labeling and record Cal Due Dates
Model Number
Test
Expected
Results
Limits
Ohms
Ohms
<.2
µ
µ
A
A
USA
300
500
300
1000
300
1000
300
500
100
300
300
300
300
100
SP
µ
ECG
A
O2
10
50
50
50
50
10
SP
µ
ECG
A
O2
50
50
µ
µ
A
A
100
500
500
500
500
100
µ
µ
A
A
5000
5000
Dash 3000/4000/5000
Serial Number
Test Equipment List
Description
Done
I
P
F
N
A
A
S
C
P
S
I
T
M
M
S
L
Monitor Power Up - (POST)
ECG 5 LD
ECG 12SL
Respiration
Temperature 1 and 2
Cardiac Output
Invasive BP
Pulse Oximetry
NBP
Analog Out, Defibrillator
Synchronization
End-Tidal CO2
Batteries
Battery A
Battery B
Printer
Display
Speaker
Network
Remote Control
BISx
Wireless LAN
Dash Port 2 Docking
Station
TRAM-RAC 2A w/ICG
(option)
TRAM-RAC 2A w/SAM
(option)
Abbreviations
DUT
Standard (S)
Option (O)
INST
CM
PM
CONF
Device Test Results
Pass ( )
Comments
Serial Number
Cal Due Date
(S)
DUT
(O)
C
O
N
N
S
Functional Checks
F
T
(S)
(option)
(S)
(S)
(option)
(option)
(S)
(S)
(S)
(option)
(option)
(option)
(option)
(S)
(S)
(option)
(option)
(option)
(option)
(option)
Device Under Test
Standard configuration test required
Test as required if the option is installed
Installation
Corrective Maintenance
Preventive Maintenance
Configuration – Standard (S) or Option (O)
Tested By
Fail ( )
Done
I
P
F
A
A
C
P
S
I
M
M
S
L
Date
2000966-542D

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

Dash 4000Dash 5000001c 051d 003a

Table of Contents