Replace Nbp Pump Assembly - GE Dash 3000 Service Manual

Patient monitor
Hide thumbs Also See for Dash 3000:
Table of Contents

Advertisement

Replace NBP pump assembly

7-40
Field replaceable units: Replace main unit parts
1. Remove the DAS assembly according to steps starting on page 7-37.
2. Remove the four screws and remove the NBP assembly from the
cover of the DAS assembly.
3. Disconnect the tube from the inner solenoid going into the NBP
assembly.
Disconnect this end of the tube.
4. Using a small, flat-blade screwdriver, remove the connector from the
NBP assembly to the DAS assembly.
Dash 3000/4000/5000
Remove four
screws.
DAS assembly FRU
NBP pump assembly FRU
2000966-542D
570A
527A
526A