GE Dash 3000 Service Manual page 193

Patient monitor
Hide thumbs Also See for Dash 3000:
Table of Contents

Advertisement

2000966-542D
Field replaceable units: Replace main unit parts
9. Carefully pull the CPU/battery housing assembly from the unit.
10. Remove the plastic connector panel from the three connectors at the
back of the CPU/battery housing assembly. Install this panel on the
new CPU/battery housing assembly before you slide the new
assembly into the rear housing.
11. Install the new assembly in the unit.
NOTE
When securing the thumb screw on the writer flex, first finger-
tighten the thumb screw. Then, use a flat-head screw driver to
tighten the screw an additional 1/4 turn.
12. If present, install the wireless card into the new CPU/battery
housing assembly.
CAUTION
Do not pull on the 802.11a/b/g antennas or use force
when connecting to the wireless card. Handle the
antennas and connectors carefully as they can easily be
damaged.
Dash 3000/4000/5000
plastic connector panel
547A
572A
7-47