Acceptance criteria
2000966-542D
Functional and electrical safety checks: Electrical safety tests
2. Power on device under test.
3. Read and record leakage current indicated on the tester.
4. Change leakage tester switches to:
Polarity – REVERSED
Neutral – CLOSED
GND (Earth) – CLOSED
5. Read and record the current leakage indicated on the tester.
6. Set the power switch on the device to OFF.
7. Repeat the steps in this procedure using the appropriate SPO2 Test
Body. Connect the SPO2 Test Body to the blue SPO2 connector of the
device under test.
If measured reading is greater than the appropriate specification below,
the device under test fails. Contact GE Technical Support.
50 µA at 120-240 VAC using the ECG cable.
Dash 3000/4000/5000
8-15