2000966-542D
Field replaceable units: Replace display assembly parts
d.
Use a 5 mm nut driver to remove the
four stand-off screws anchoring the
display to the rubber display isolator.
NOTE
The stand-offs are shimmed at the
factory. Save the washers and make
sure you put them in the same
location when you reassemble the
display
e.
Lift out the LCD display from the
rubber display isolator and set aside.
3. Dampen a clean, soft cloth with window cleaner and carefully clean
and dry the LCD display and the display filter to remove fingerprints
and dust particles.
4. If required, replace the display filter, front bezel, and insert.
5. Install the existing LCD display and reassemble the display
assembly in reverse order.
6. Complete the procedures in
Dash 3000/4000/5000
886A
887A
"Recommended checkout"
on page 7-52.