GE DMS Go Operator's Manual page 59

Table of Contents

Advertisement

2.3.2e
Setting the Zero Block Configuration (CONFIG-ZERO BLOCK)
If you are using a zero block for your probe, you can enter its thickness and velocity configuration into the DMS Go. (See "Setting the Probe
Zero (PROBE & CAL - ZERO)" on page 63.)
In the MAIN menu, scroll to the CONFIG submenu using the joystick (
1.
2.
Use the joystick (
(Figure 11 below).
3.
Move the joystick (
) to change the unit (inches or mm), thickness (in inches or mm) or the velocity (in inches/microseconds or m/s)
to the value engraved on the GE-supplied reference block. These values will always be displayed in metric (mm). Press F2 (DONE) to
confirm the entry, or F1 (EXIT) to close the window without changing the configuration.
4.
After making your choice, use the joystick (
DMS Go Operator's Manual
Your Complete Source for
Testing Equipment. Since 1969!
) to select the ZERO BLOCK function, then move the joystick (
Figure 11: Zero Block Configuration Window
) to move to another function, or press HOME to return to the Main Menu.