Connecting
7.2 Gas connections
Installation of FKM hoses:
● Connect the PTFE pipe sections (30 mm long) to the screwed glands.
● Slide the FKM hoses onto the PTFE pipes.
● Secure the hoses with hose clamps.
7.2.3

ULTRAMAT 7

Arrangement of the gas connections
P
1
2
3
4
Figure 7-9
Note
Identifier of the gas connections
The identifier of the gas connections is imprinted as numbers 1 to 4 on the module plate.
The sample gas connections are made of stainless steel, material no. 1.4404 or Hastelloy,
material no. 2.4819. The reference gas connection is made of stainless steel, mat. no. 1.4404.
The gas connections are provided as connection fittings with a pipe diameter of 6 mm.
96
1
ULTRAMAT 7
Atmospheric pressure sensor
Sample gas inlet
Sample gas outlet
Reference gas outlet
Reference gas inlet
ULTRAMAT 7 gas connections
Rack-mounted device
Operating Instructions, 05/2015, A5E31874006-03