Table 3-6 Plug & Measure - Installation/Replacement Of A Pm; Table 3-7 Plug & Measure - Replacement Of A Pu/Installation Of A Used Am - Siemens SIPROCESS GA700 Operating Instructions Manual

Continuous gas analysis, rack-mounted device
Hide thumbs Also See for SIPROCESS GA700:
Table of Contents

Advertisement

Description
3.7 Combined operation
Case
PM
Un‐
change
d
Un‐
change
d
Un‐
change
d
Table 3-6
Case
PM
New
New
Serial number stored in the PU is also the serial number of the device
1)
Table 3-7
Case
PM
Re‐
placed
Re‐
placed
54
Description
AMx
Re‐
An AM is replaced by an AM of a differ‐
placed
ent type.
New
An additional AM of the same type is to
be installed.
An additional AM of a previously instal‐
led type is installed again.
New
An additional AM of a type that has not
been connected before is to be installed.
Plug & Measure - Installation/replacement of a PM
Description
AMx
Un‐
The PM electronics is replaced. Quota‐
change
tion marks are displayed instead of the
d
serial numbers. At least one AM re‐
mains in the device.
Un‐
PM electronics with new serial number
change
is installed.
d
Plug & Measure - Replacement of a PU/Installation of a used AM
Description
AMx
Re‐
Assembly from inventory: Each module
placed
has a serial number and has been con‐
nected to another module before. AM of
the same type is replaced.
Re‐
Assembly from inventory: Each module
placed
has a serial number and has been con‐
nected to another module before. AM of
a different type is replaced.
Data synchronization/Action
Automatic data synchronization, no ac‐
tion required by user.
User action. Confirm <ENTER> or re‐
ject <ESC>.
The data is transmitted after confirma‐
tion, except the identification data and
the calibration parameters.
Automatic data synchronization, no ac‐
tion required by user.
Data synchronization/Action
Automatic data synchronization, no ac‐
tion required by user.
When there are two AMs in the device,
the new PM always synchronizes with
AM1.
The PM receives the data of the previ‐
ously used PM, including its serial num‐
ber
1)
Automatic data synchronization, no ac‐
tion required by user.
In this scenario, the replacement of the
PM electronics results in a loss of data
of the previously installed PM. In addi‐
tion, the serial number is overwritten.
Data synchronization/Action
User action. Confirm with <ENTER> or
reject with <ESC>.
When the process is confirmed, stored
AM data is transmitted to the replaced
AM.
Automatic data synchronization, no ac‐
tion required by user. Each module re‐
tains its data.
Rack-mounted device
Operating Instructions, 05/2015, A5E31874006-03