10.4

Replacing the processing unit

Procedure
1. Isolate the device from power.
2. Remove the enclosure cover. To do so, unscrew the two nuts on the rear panel of the device
and push the enclosure cover to the back.
Figure 10-1
3. Loosen the two screws and remove the printed circuit board holder
4. Disconnect the following cables depending on the installation situation:
– Power cable
– 26-pin ribbon cable to display
– Ethernet cable
– 12-pin ribbon cable to analyzer modules (AM) and option module OM 2.2
– 50-pin ribbon cable to option module 1.1
5. Loosen the screws (outer, rear) and remove the processing unit (PU)
6. Install the new PU in the reverse order.
Tighten the screws on the rear panel with a torque of 1.3 Nm.
Tighten the screws of the PCB holder on the power supply, and the cap nuts on the
enclosure cover with a torque of 2.5 Nm.
Rack-mounted device
Operating Instructions, 05/2015, A5E31874006-03
Power cable
Ethernet cable
12-pin ribbon cable (CAN bus)
PU and screws
26-pin ribbon cable (display)
PCB holder and screws
50-pin ribbon cable
Connection cables for processing unit
10.4 Replacing the processing unit
Maintenance and service
.
.
119