Maintenance and service

10.3 Replace basic device

10.3
Replace basic device
Requirement
The basic device is completely parameterized, the serial number and factory data are saved.
The user-specific parameters are stored externally, e.g. with SIMATIC PDM.
Procedure for removal
1. Shut off all gas lines to the device.
2. Ensure that no hazardous gases escape when removing the gas lines to the device.
3. To remove the analyzer modules, follow the procedure described in section Removing an
analyzer module (Page 75).
4. To remove the option modules, follow the procedure described in section Installing option
modules (Page 76).
5. Dismantle the basic device.
Procedure for removal
1. Mount the new basic device
2. Install the previously used analyzer module (AM) into the new basic device.
Proceed as described in section Installation of the analyzer module (Page 71).
– Subsequent to switching on the device, the AM is recognized as new.
– Each AM retains its parameter assignment.
3. To install the option modules again, follow the procedure described in sections Installing
option modules (Page 77) and Wiring option modules (Page 78).
4. To commission the analyzer modules, follow the procedure described in section
Commissioning (Page 101).
5. Enter the user-specific parameters of the basic device again.
If the parameters have been saved on a PC using SIMATIC PDM, download these to the
device.
See also
References (Page 141)
118
Also read the information in the Plug & Measure section of the operating manuals.
Rack-mounted device
Operating Instructions, 05/2015, A5E31874006-03