Philips M3150B Installation And Service Manual page 460

Information center system
Table of Contents

Advertisement

Configuring the System
Table 8-7. M3185 Clinical Network Components - Test and Inspection Matrix
Test Block
Test or Inspection to Perform
Name
Performance:
Perform an operational test of the Clinical Network by
executing a data passing operation from each connected
Information Center to every other Information Center,
Client, or Printer on the Network
For each M3150 and M3155 Information Center:
Verify that a review application (e.g. Wave Review) can
be executed. This verifies connection to the M3154
Database Server (if a M3155 Information Center).
For each M3151 Client:
Verify that waveforms from every Information Center on
the Network having a connection to an SDN can be
viewed on the Client.
Verify that a review application (e.g. Wave Review) can
be executed. This verifies connection to the M3154
Database Server
For each wireless Patient Monitor:
Verify that waveforms from each wireless patient monitor
on the network are displayed on an Information Center
and that the waveforms are continuous.
Safety
No safety test is required
8-18
Start-up and Configuration
Expected Results
Expected answers are
"Yes". If so,
Performance test passed.
What to Record
on Service
Record
P:P or P:F
where P = Pass
F = Fail
S:NA
where NA = Not
required

Advertisement

Table of Contents
loading

This manual is also suitable for:

M3170M3155bM3151bIntellivue seriesM3169aM3154b

Table of Contents