Site Preparation
4-9, consult Table 4-3, Space and Weight Requirements for Information Center Installations and/
or the documentation for the units that comprise the system.
Environmental
The environment where the Information Center system will be used should be reasonably free
Requirements
from vibration, dust, and corrosive or explosive gases. The ranges of ambient operating
conditions -- temperature and relative humidity (RH) -- and the maximum heat dissipated for the
Information Center installations shown in Figure 4-9 are given in the following table. To
determine similar specifications for Information Center installations different from those in
Figure 4-9, consult Table 4-4 and/or the documentation for the units that comprise the system.
Caution
If the environmental specifications of the Information Center system are not met, damage
may be caused to the equipment.
System
4-22
Site Planning and Preparation
Table 4-3 Space and Weight Requirements for Information Center
Height
System
cm (in.)
A
42.2 (16.6)
B
42.2 (16.6)
C
42.2 (16.6)
Table 4-4. Environmental Requirements for Information Center Installations
Heat (max)
Temperature
BTU/hr.
A
2928
15 - 30
59 - 86
B
3270
15 - 30
59 - 86
C
2928
15 - 30
59 - 86
Installations
Width
Depth
cm (in.)
cm (in.)
72 (28)
64 (25)
114 (45)
64 (25)
89 (35)
66 (26)
Relative
Humidity
o
C
20 - 80%
o
o
o
F
@ 30
C (86
F)
o
C
20 - 80%
o
o
o
F
@ 30
C (86
F)
o
C
20 - 80%
o
o
o
F
@ 30
C (86
F)
Weight
kg. (lbs.)
19.7 (43.3)
36.7 (80.7)
34.7 (76.3)
*
Includes PC and UPS
+1 for optional LaserJet printer
Notes
Assumes single, medium display, PC,
Recorder, and UPS.
Assumes dual, medium displays, PC,
Recorder, and UPS.
Assumes single, medium display, PC,
Recorder, and UPS.
# Power
*
Outlets
2