No. 2D730-148E*O
3.16
Removing the MO
(1)
Removing the cables
(a)
Remove the two rubber caps at the bottom of the left side cover.
(b)
Remove the two M4 screws <A> to remove the left side cover.
(c)
Remove the four M3 screws <B> to remove the internal cover.
(d)
Remove the internal SCSI cable (J505) and power cable (J502 or J503) that are
connected to the MO.
(2)
Removing the MO
(a)
Remove the blank cover.
(b)
Remove the two M3 screws CC> to extract the MO
Left side cover
:R)
Cap
I
Screw CA>
(SBPP4-12SCR)
3-50